BodyTrace BP Monitor Instructions
Installing the Batteries
- Slide open the battery door on the back of the device.
- Follow the diagram inside the device to install the batteries provided with the device.
- Slide the battery door closed.
Securing the Cuff
- Plug the connector on cuff tube to device.
- Expose your left upper arm by removing or adjusting clothing and jewelry.
- Make sure blood flow is not constricted by a rolled up sleeve.
- Open the cuff and loosen fully.
- Orient the cuff so that the tube exits towards the hand.
- Place your arm through the cuff loop, with your palm facing up.
- Position the cuff's edge about an inch (2-3cm) above the elbow.
- Align the ะค marker (located to the right of tube exit) with the center of your arm.
Taking a Measurement
- Lay your arm on a table with your palm facing up.
- Sit up straight with feet flat on the ground.
- The cuff should be even with your heart.
- Press the START/STOP button to turn on.
- Wait while the device takes the measurement.
- The readings will appear on the device and cuff will deflate.
- The device automatically transmits your results to your provider.
- When the measurement is successfully transmitted, the device will display the image displayed below.
Error Codes
- Er1 - low battery; replace four AA batteries
- Er2 - movement detected; retake reading
- Er3 - Measurement out of range; check cuff is secure and retake reading. If unwell, contact medical professional
- Er4, Er5, Er6 - no cellular coverage; move to another area closer to a window
- Er7 or higher - contact GatherMed